Model:FT-S12
Brand:fengtu
1、Dissolved Ozone Sensor Product Overview
Dissolved Ozone Sensor designed for continuous, online monitoring of ozone concentration in water; requires no sample pretreatment or colorimetric reagents.Dissolved Ozone Sensor is a three-electrode potentiostatic sensor developed by our company, featuring proprietary intellectual property rights. It is designed to measure the concentration of disinfectants such as ozone and is primarily used in applications including treated water at water treatment plants, water distribution networks, secondary water supply systems, end-of-line water supply points, and swimming pools. The sensor incorporates an ARM processor and high-efficiency filtering algorithms to effectively mitigate noise interference. It is equipped with an RS-485 interface, facilitating seamless integration into computer-based or networked monitoring systems.
2、Dissolved Ozone Sensor Measurement Principle
The constant-potential sensor employs a three-electrode measurement system consisting of two platinum electrodes and one silver/silver chloride electrode. It incorporates a high-precision potentiostat that maintains a stable potential at the working electrode. Disinfectant components—such as hypochlorous acid—undergo redox reactions at the working electrode, generating a current that follows Faraday’s law, which allows for the determination of the disinfectant concentration.
3、Dissolved Ozone Sensor Product Features
1. The electrochemical sensor consumes no reagents and produces no pollutant emissions.
2. The membrane-free design eliminates the need for membrane replacement or the addition of electrolyte.
3. A three-electrode configuration ensures a stable zero point and high sensitivity.
4. An integrated high-precision sampling circuit provides excellent linearity.
5. When pH levels remain relatively stable, the sensor’s readings can be accurately compared with those obtained via the DPD measurement method.
4、Dissolved Ozone Sensor Technical Specifications
The technical specifications of the product are shown in Table 4.1.
| System | Power supply | DC 9-28V |
| Power | 0.5W (peak power) | |
| Measurement parameters | Ozone, temperature | |
| Communication Interface | RS485+4-20mA | |
| Electrode material | Platinum, silver chloride | |
| Design life | >18 months | |
| Water sample flow rate | >400mL/min, and the flow rate must be kept stable | |
| Operating temperature | -5-45℃ (non-freezing) | |
| Withstand pressure | <0.1MPa | |
| Ozone | Measurement Methods | Three-electrode amperometric method |
| Range | 0-5/20mg/L | |
| Resolution | 0.001mg/L | |
| Detection Limit | 0.030mg/L or less | |
| Accuracy | ±3% (DPD comparison error ±10% or ±0.05 mg/L, whichever is greater) | |
| Repetitive | ±3% | |
| Response Time (T90) | ≤90 seconds | |
| Calibration Methods | Single-point calibration, two-point calibration | |
| Sample pH range | 4-9 | |
| Temperature | Measurement Methods | Thermistor method |
| Range | -5-60℃ | |
| Resolution | 0.1℃ | |
| Accuracy | ±0.4℃ | |
| Repetitive | 0.2℃ | |
| Response Time (T90) | ≤25 seconds | |
| Calibration Methods | Single-point calibration |
Dissolved Ozone Sensor Product Dimensions Drawing

Carbon monoxide and nitrogen dioxide gases in mines pose great hazards to the safety of miners' lives and mine operations, which is the main reason for their monitoring.Carbon monoxide is a colorless, odorless, and tasteless gas. Its affinity for human hemoglobin is 200 - 300 times higher than t...
Meteorological monitoring technologies are becoming increasingly indispensable in contemporary society with the increasing emphasis on rural development, agricultural progress and ecological protection. Especially in the environmental monitoring of rural streets, the integrated ultrasonic sensing te...
light sensors measure illumination, respond to changes in the amount of light received, detect light and convert it into electrical signals for both indoor and outdoor use....
Photovoltaic IV curve testing is a crucial means to evaluate the electrical performance of photovoltaic modules/strings, and the Portable IV Curve Tester is an important tool to achieve this testing.The Portable IV Curve Tester is specifically used to measure the current - voltage relationship of ph...